Performance of infrared ear and forehead thermometers: a comparative study in 205 febrile and afebrile children

Author

Listed:
  • Patricia A Hamilton
  • Lorenzo S Marcos
  • Michelle Secic

Abstract

Aims and objectives This study compared readings from two professional‐grade, commercially available infrared (IR) thermometers, the ThermoScan® PRO 4000 prewarmed tip ear thermometer and the Temporal Scanner™ TAT‐5000 temporal artery thermometer. Background The repeatability and precision of readings from IR thermometers for professional use were questioned in the past, but in recent years, these types of thermometers have been technologically improved, so their ability to replicate standard temperature readings reliably should be re‐examined. Design Febrile and afebrile children were recruited from the emergency department, overflow treatment areas and the paediatric intensive care unit of a large hospital in Argentina. Each child had a randomised sequence of seven temperature readings, including three from the ear, three from the forehead or behind the ear and one reference oral or rectal reading. Methods Temperature readings were taken with the ThermoScan PRO 4000, the Temporal Scanner TAT‐5000 and the monitor mode of SureTemp® Plus, a widely used professional‐grade contact thermometer, for reference. Results Of 205 children, 46% were febrile, per reference thermometer readings. While mean ThermoScan PRO 4000 febrile measurements did not differ significantly from reference, mean Temporal Scanner TAT‐5000 febrile measurements were significantly lower (by a mean of 0·42 °C) than the reference. Overall bias of the ThermoScan PRO 4000 was significantly lower than that of the Temporal Scanner TAT‐5000; repeatability was 1·5 times higher, and overall false‐negative rate was about a third that of Temporal Scanner TAT‐5000, when compared to the reference. Conclusions This study indicates that the ThermoScan PRO 4000 provides measurements closer to those of a professional‐grade contact thermometer when compared to the Temporal Scanner TAT‐5000. Relevance to clinical practice The odds of misclassifying a febrile child as non‐febrile were about threefold higher with the Temporal Scanner TAT‐5000 temporal artery thermometer than with the ThermoScan® PRO 4000 ear thermometer.

Suggested Citation

  • Patricia A Hamilton & Lorenzo S Marcos & Michelle Secic, 2013. "Performance of infrared ear and forehead thermometers: a comparative study in 205 febrile and afebrile children," Journal of Clinical Nursing, John Wiley & Sons, vol. 22(17-18), pages 2509-2518, September.
  • Handle: RePEc:wly:jocnur:v:22:y:2013:i:17-18:p:2509-2518
    DOI: 10.1111/jocn.12060
    as

    Download full text from publisher

    File URL: https://doi.org/10.1111/jocn.12060
    Download Restriction: no

    File URL: https://libkey.io/10.1111/jocn.12060?utm_source=ideas
    LibKey link: if access is restricted and if your library uses this service, LibKey will redirect you to where you can use your library subscription to access this item
    ---><---

    More about this item

    Statistics

    Access and download statistics

    Corrections

    All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:wly:jocnur:v:22:y:2013:i:17-18:p:2509-2518. See general information about how to correct material in RePEc.

    If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.

    We have no bibliographic references for this item. You can help adding them by using this form .

    If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.

    For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: Wiley Content Delivery (email available below). General contact details of provider: https://doi.org/10.1111/(ISSN)1365-2702 .

    Please note that corrections may take a couple of weeks to filter through the various RePEc services.

    IDEAS is a RePEc service. RePEc uses bibliographic data supplied by the respective publishers.