Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients

Author

Listed:
  • Richard A Parker
  • Christopher J Weir
  • Noah Rubio
  • Roberto Rabinovich
  • Hilary Pinnock
  • Janet Hanley
  • Lucy McCloughan
  • Ellen M Drost
  • Leandro C Mantoani
  • William MacNee
  • Brian McKinstry

Abstract

Introduction: The Bland-Altman limits of agreement method is widely used to assess how well the measurements produced by two raters, devices or systems agree with each other. However, mixed effects versions of the method which take into account multiple sources of variability are less well described in the literature. We address the practical challenges of applying mixed effects limits of agreement to the comparison of several devices to measure respiratory rate in patients with chronic obstructive pulmonary disease (COPD). Methods: Respiratory rate was measured in 21 people with a range of severity of COPD. Participants were asked to perform eleven different activities representative of daily life during a laboratory-based standardised protocol of 57 minutes. A mixed effects limits of agreement method was used to assess the agreement of five commercially available monitors (Camera, Photoplethysmography (PPG), Impedance, Accelerometer, and Chest-band) with the current gold standard device for measuring respiratory rate. Results: Results produced using mixed effects limits of agreement were compared to results from a fixed effects method based on analysis of variance (ANOVA) and were found to be similar. The Accelerometer and Chest-band devices produced the narrowest limits of agreement (-8.63 to 4.27 and -9.99 to 6.80 respectively) with mean bias -2.18 and -1.60 breaths per minute. These devices also had the lowest within-participant and overall standard deviations (3.23 and 3.29 for Accelerometer and 4.17 and 4.28 for Chest-band respectively). Conclusions: The mixed effects limits of agreement analysis enabled us to answer the question of which devices showed the strongest agreement with the gold standard device with respect to measuring respiratory rates. In particular, the estimated within-participant and overall standard deviations of the differences, which are easily obtainable from the mixed effects model results, gave a clear indication that the Accelerometer and Chest-band devices performed best.

Suggested Citation

  • Richard A Parker & Christopher J Weir & Noah Rubio & Roberto Rabinovich & Hilary Pinnock & Janet Hanley & Lucy McCloughan & Ellen M Drost & Leandro C Mantoani & William MacNee & Brian McKinstry, 2016. "Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients," PLOS ONE, Public Library of Science, vol. 11(12), pages 1-15, December.
  • Handle: RePEc:plo:pone00:0168321
    DOI: 10.1371/journal.pone.0168321
    as

    Download full text from publisher

    File URL: https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0168321
    Download Restriction: no

    File URL: https://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0168321&type=printable
    Download Restriction: no

    File URL: https://libkey.io/10.1371/journal.pone.0168321?utm_source=ideas
    LibKey link: if access is restricted and if your library uses this service, LibKey will redirect you to where you can use your library subscription to access this item
    ---><---

    References listed on IDEAS

    as
    1. Carstensen Bendix & Simpson Julie & Gurrin Lyle C, 2008. "Statistical Models for Assessing Agreement in Method Comparison Studies with Replicate Measurements," The International Journal of Biostatistics, De Gruyter, vol. 4(1), pages 1-28, July.
    Full references (including those not matched with items on IDEAS)

    Citations

    Citations are extracted by the CitEc Project, subscribe to its RSS feed for this item.
    as


    Cited by:

    1. Trang Nguyen Phuc Thu & Alfredo I Hernández & Nathalie Costet & Hugues Patural & Vincent Pichot & Guy Carrault & Alain Beuchée, 2019. "Improving methodology in heart rate variability analysis for the premature infants: Impact of the time length," PLOS ONE, Public Library of Science, vol. 14(8), pages 1-14, August.

    Most related items

    These are the items that most often cite the same works as this one and are cited by the same works as this one.
    1. Liao Jason J. Z. & Capen Robert, 2011. "An Improved Bland-Altman Method for Concordance Assessment," The International Journal of Biostatistics, De Gruyter, vol. 7(1), pages 1-17, January.
    2. Kerryn A Moore & Julie A Simpson & Kyla H Thomas & Marcus J Rijken & Lisa J White & Saw Lu Moo Dwell & Moo Kho Paw & Jacher Wiladphaingern & Sasithon Pukrittayakamee & François Nosten & Freya J I Fowk, 2015. "Estimating Gestational Age in Late Presenters to Antenatal Care in a Resource-Limited Setting on the Thai-Myanmar Border," PLOS ONE, Public Library of Science, vol. 10(6), pages 1-17, June.

    More about this item

    Statistics

    Access and download statistics

    Corrections

    All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:plo:pone00:0168321. See general information about how to correct material in RePEc.

    If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.

    If CitEc recognized a bibliographic reference but did not link an item in RePEc to it, you can help with this form .

    If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.

    For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: plosone (email available below). General contact details of provider: https://journals.plos.org/plosone/ .

    Please note that corrections may take a couple of weeks to filter through the various RePEc services.

    IDEAS is a RePEc service. RePEc uses bibliographic data supplied by the respective publishers.