Author
Listed:
- Mohsen Bakouri
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia
Department of Physics, College of Arts, Fezzan University, Traghen City 71340, Libya)
- Ahmad Alassaf
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia)
- Khaled Alshareef
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia)
- Amor Smida
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia
Microwave Electronics Research Laboratory, Department of Physics, Faculty of Mathematical, Physical and Natural Sciences of Tunis, Tunis ElManar University, Tunis 2092, Tunisia)
- Ibrahim AlMohimeed
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia)
- Abdulrahman Alqahtani
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia
Department of Medical Equipment Technology, College of Applied Medical Sciences, Prince Sattam Bin Abdulaziz University, Al-Kharj 16278, Saudi Arabia)
- Mohamed Abdelkader Aboamer
(Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Al-Majmaah 11952, Saudi Arabia)
- Yousef Alharbi
(Department of Medical Equipment Technology, College of Applied Medical Sciences, Prince Sattam Bin Abdulaziz University, Al-Kharj 16278, Saudi Arabia)
Abstract
Installing and developing a sophisticated control system to optimize left ventricular assist device (LVAD) pump speed to meet changes in metabolic demand is essential for advancing LVAD technology. This paper aims to design and implement a physiological control method for LVAD pumps to provide optimal cardiac output. The method is designed to adjust the pump speed by regulating the pump flow based on a predefined set point (operating point). The Frank–Starling mechanism technique was adopted to control the set point within a safe operating zone (green square), and it mimics the physiological demand of the patient. This zone is predefined by preload control lines, which are known as preload lines. A proportional–integral (PI) controller was utilized to control the operating point within safe limits to prevent suction or overperfusion. In addition, a PI type 1 fuzzy logic controller was designed and implemented to drive the LVAD pump. To evaluate the design method, rest, moderate, and exercise scenarios of heart failure (HF) were simulated by varying the hemodynamic parameters in one cardiac cycle. This evaluation was conducted using a lumped parameter model of the cardiovascular system (CVS). The results demonstrated that the proposed control method efficiently drives an LVAD pump under accepted clinical conditions. In both scenarios, the left ventricle pressure recorded 112 mmHg for rest and 55 mmHg for exercise, and the systematic flow recorded 5.5 L/min for rest and 1.75 L/min for exercise.
Suggested Citation
Mohsen Bakouri & Ahmad Alassaf & Khaled Alshareef & Amor Smida & Ibrahim AlMohimeed & Abdulrahman Alqahtani & Mohamed Abdelkader Aboamer & Yousef Alharbi, 2022.
"A Feasible Method to Control Left Ventricular Assist Devices for Heart Failure Patients: A Numerical Study,"
Mathematics, MDPI, vol. 10(13), pages 1-17, June.
Handle:
RePEc:gam:jmathe:v:10:y:2022:i:13:p:2251-:d:848995
Download full text from publisher
Corrections
All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:gam:jmathe:v:10:y:2022:i:13:p:2251-:d:848995. See general information about how to correct material in RePEc.
If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.
If CitEc recognized a bibliographic reference but did not link an item in RePEc to it, you can help with this form .
If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.
For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: MDPI Indexing Manager (email available below). General contact details of provider: https://www.mdpi.com .
Please note that corrections may take a couple of weeks to filter through