IDEAS home Printed from https://ideas.repec.org/a/gam/jijerp/v12y2015i6p5886-5904d50212.html
   My bibliography  Save this article

Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board

Author

Listed:
  • Tobias Seckler

    (The Research Center for Bioelectromagnetic Interaction (FEMU), Institute and Out-patient Clinic of Occupational Medicine, RWTH Aachen University, Pauwelsstr 30, 52074 Aachen, Germany)

  • Kai Jagielski

    (The Research Center for Bioelectromagnetic Interaction (FEMU), Institute and Out-patient Clinic of Occupational Medicine, RWTH Aachen University, Pauwelsstr 30, 52074 Aachen, Germany)

  • Dominik Stunder

    (The Research Center for Bioelectromagnetic Interaction (FEMU), Institute and Out-patient Clinic of Occupational Medicine, RWTH Aachen University, Pauwelsstr 30, 52074 Aachen, Germany)

Abstract

Electromagnetic interference is a concern for people wearing cardiovascular implantable electronic devices (CIEDs). The aim of this study was to assess the electromagnetic compatibility between CIEDs and the magnetic field of a common wireless charging technology. To do so the voltage induced in CIEDs by Qi A13 design magnetic fields were measured and compared with the performance limits set by ISO 14117. In order to carry this out a measuring circuit was developed which can be connected with unipolar or bipolar pacemaker leads. The measuring system was positioned at the four most common implantation sites in a torso phantom filled with physiological saline solution. The phantom was exposed by using Helmholtz coils from 5 µT to 27 µT with 111 kHz sine‑bursts or by using a Qi A13 design wireless charging board (Qi‑A13‑Board) in two operating modes “power transfer” and “pinging”. With the Helmholtz coils the lowest magnetic flux density at which the performance limit was exceeded is 11 µT. With the Qi‑A13‑Board in power transfer mode 10.8% and in pinging mode 45.7% (2.2% at 10 cm distance) of the performance limit were reached at maximum. In neither of the scrutinized cases, did the voltage induced by the Qi‑A13‑Board exceed the performance limits.

Suggested Citation

  • Tobias Seckler & Kai Jagielski & Dominik Stunder, 2015. "Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board," IJERPH, MDPI, vol. 12(6), pages 1-19, May.
  • Handle: RePEc:gam:jijerp:v:12:y:2015:i:6:p:5886-5904:d:50212
    as

    Download full text from publisher

    File URL: https://www.mdpi.com/1660-4601/12/6/5886/pdf
    Download Restriction: no

    File URL: https://www.mdpi.com/1660-4601/12/6/5886/
    Download Restriction: no
    ---><---

    Citations

    Citations are extracted by the CitEc Project, subscribe to its RSS feed for this item.
    as


    Cited by:

    1. Feng Wen & Xueliang Huang, 2017. "Human Exposure to Electromagnetic Fields from Parallel Wireless Power Transfer Systems," IJERPH, MDPI, vol. 14(2), pages 1-15, February.

    More about this item

    Keywords

    intermediate frequency magnetic fields; pacemaker; defibrillator; electromagnetic interference; wireless charging; wireless power transfer; Qi A13 design; performance limits; EMF risk assessment;
    All these keywords.

    JEL classification:

    • A13 - General Economics and Teaching - - General Economics - - - Relation of Economics to Social Values

    Statistics

    Access and download statistics

    Corrections

    All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:gam:jijerp:v:12:y:2015:i:6:p:5886-5904:d:50212. See general information about how to correct material in RePEc.

    If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.

    We have no bibliographic references for this item. You can help adding them by using this form .

    If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.

    For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: MDPI Indexing Manager (email available below). General contact details of provider: https://www.mdpi.com .

    Please note that corrections may take a couple of weeks to filter through the various RePEc services.

    IDEAS is a RePEc service. RePEc uses bibliographic data supplied by the respective publishers.